
Abstract
Objectives
Cardiopulmonary bypass (CPB) and extracorporeal membrane oxygenation (ECMO) or life support (ECLS) circuits are built from polymers and might release polymeric micro- and nanoparticles (MNP) into the circulation. MNPs seem to provoke inflammation, oxidative stress, and apoptosis, which are also side effects of extracorporeal circulation. Thus, we investigated the MNP release from CPB and ECMO/ECLS circuits.
Methods
A CPB and ECMO/ECLS circuit was filled with saline solution, and circulation was initiated for 5 h and 7 d, respectively. Samples were taken from both circuits and filtered through a silicon membrane. MNPs were detected and quantified using optical microscopy and micro Raman spectroscopy.
Results
During circulation, polyvinyl chloride (PVC) and polymethyl methacrylate (PMMA) were detected in the CPB perfusate. After 5 h of circulation, polyethylene terephthalate (PET) was detected. In the ECMO/ECLS circuit, time-dependent accumulation of polymeric fragments was detected. Finally, particles of polyethylene (PE), polystyrene (PS), PET, and PVC were identified. The particle-size distribution extended from initially <2 µm to finally >10 µm with increasing circulation time.
Conclusions
CPB and ECMO/ECLS circuits release MNPs. The number of MNPs increases over the period of use. A larger number of circuits and of health effects of identified MNPs, should be investigated.